(a) Start up.
(1) Turn off ultraviolet sterilizer if so equipped.
(2) Turn on all blowers and cabinet illumination lights.
(3) Allow 5 minutes of operation to purge system, check flow alarm system audio and visual alarm function if so equipped.
(4) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.
(b) Shut down.
(1) Decontaminate and remove all items from interior work area.
(2) Decontaminate readily accessible interior surfaces with a disinfectant appropriate for the agents or suspected agents present.
(3) Turn on ultraviolet sterilizer if so equipped.
(4) Allow 5 minutes of operation to purge system.
(5) Turn off cabinet blower.
NOTE: Authority cited: Section 142.3, Labor Code. Reference: Section 142.3, Labor Code.